医用炭/炭复合材料人工髋关节的磨损机理研究

展开
  • (西北工业大学 凝固技术国家重点实验室, 西安710072)

收稿日期: 2010-03-04

  修回日期: 2010-05-12

  网络出版日期: 2010-08-25

基金资助

“111” Project of China (B08040); National Natural Science Foundation of China(50832004)

Wear Mechanism of Biomedical Carbon/Carbon Composites for Artificial Hip Joints

Expand
  • (State Key Laboratory of Solidification Processing, Northwestern Polytechnical University, Xi’an 710072, China)

Received date: 2010-03-04

  Revised date: 2010-05-12

  Online published: 2010-08-25

摘要

髋关节模拟试验机在多方性复合运动方式、恒定载荷和血清润滑条件模拟了炭/炭复合材料人工髋关节在临床应用中的摩擦学行为. 采用扫描电镜、能谱和激光粒度分析仪分析了其磨损表面状态和磨损颗粒特性. 结果表明, 医用炭/炭复合材料人工髋关节在不同碳纤维方向呈现不同的磨损特性, 其磨损颗粒具有四种形貌并且尺寸分布在亚微米至数十微米之间, 其磨损机理主要是微切削作用和磨粒磨损.

本文引用格式

李贺军, 张磊磊, 卢锦花, 李克智, 付前刚, 赵雪妮, 曹 盛 . 医用炭/炭复合材料人工髋关节的磨损机理研究[J]. 无机材料学报, 2010 , 25(9) : 999 -1002 . DOI: 10.3724/SP.J.1077.2010.10144

Abstract

A hip joint simulator was employed to predict the clinical wear behavior with testing conditions of a cross-path motion, a constant load and a lubricant of serum. The worn surface and the wear particles generated were analyzed by scanning electron microscope, energy dispersive spectroscope and laser particle size analyzer. The results show that the biomedical C/C composites for artificial hip joints have different wear features according to the carbon fibre orientation. Their wear particles have four kinds of morphologies with a size ranging from submicron to tens of micrometers. The wear mechanisms are mainly micro-cutting and abrasive wear.

参考文献

[1]Zhang L L, Hu T, Li H J, et al. Wear particles of carbon/carbon composite artificial hip joints.Journal of Inorganic Materials,2010, 25(4):349-353
[2]Xiong X B, Zeng X R, Zou C L, et al. Strong bonding strength between HA and (NH4)2S2O8-treated carbon/carbon composite by hydrothermal treatment and induction heating.Acta Biomater,2009, 5(5):1785-1790
[3]Pesakova V, Klezl Z, Balik K, et al. Biomechanical and biological properties of the implant material carbon-carbon composite covered with pyrolytic carbon.J. Mater. Sci.: Mater. Med,2000, 11(12):793-798
[4]Wang G H, Yu S, Zhu S H, et al. Biological properties of carbon/carbon implant composites with unique manufacturing processes.J. Mater. Sci.: Mater. Med,2009, 20(2):2487-2492
[5]Fu T, He L P, Han Y, et al. Induction of bonelike apatite on carbon-carbon composite by sodium silicate.Mater. Lett,2003, 57(22/23):3500-3503
[6]Zhang L L, Li H J, Li K Z, et al. Effect of surface roughness of carbon/carbon composites on osteoblasts growth behaviour.Journal of Inorganic Materials,2008, 23(2):341-345
[7]Zhai Y Q, Li K Z, Li H J, et al. Influence of NaF concentration on fluorine-containing hydroxyapatite coating on carbon/carbon composites.Mater. Chem. Phys,2007, 106(1):22-26
[8]Lewandoswska-Szumie M, Komender J, Gorecki A, et al. Fixation of carbon fibre-reinforced carbon composite implanted into bone.J. Mater. Sci.: Mater. Med,1997, 8(2):485-488
[9]Shi W, Dong H, Bell T. Tribological behaviour and microscopic wear mechanisms of UHMWPE sliding against thermal oxidation-treated Ti6Al4V.Mater. Sci. Eng. A,2000, 291(1/2):27-36
[10]Gonzalez-Mora V A, Hoffmann M, Stroosnijder R, et al. Wear tests in a hip joint simulator of different CoCrMo counterfaces on UHMWPE.J. Mat. Sci. Eng. C,2009, 29(3):153-158
[11]Ozcan S, Filip P. Microstructure and wear mechanisms in C/C composites.Wear,2005, 259(1-6):642-650
[12]Yen B K, Ishihara T. An investigation of friction and wear mechanisms of carbon-carbon composites in nitrogen and air at elevated temperatures.Carbon,1996, 34(4):489-498
[13]Rietsch J C, Dentzer J, Dufour A, et al. Characterizations of C/C composites and wear debris after heavy braking demands.Carbon,2009, 47(1):85-93
[14]Howling G I, Sakoda H, Antonarulrajah A, et al. Biological response to wear debris generated in carbon based composites as potential bearing surfaces for artificial hip joints.J. Biomed. Mater. Res. B,2003, 67B(2):758-763
[15]Howling G I, Ingham E, Sakoda H, et al. Carbon-carbon composite bearing materials in hip arthroplasty: analysis of wear and biological response to wear debris.J. Mater. Sci.: Mater. Med,2004, 15(1):91-98
[16]Ge S R, Wang S B, Huang X L. Increasing the wear resistance of UHMWPE acetabular cups by adding natural biocompatible particles.Wear,2009, 267(5-8):770-776

文章导航

/

〈 〉